Hyperbaric Recovery
Short hyperbaric oxygen sessions for cruise guests and resort visitors.
Schematic layout (preliminary), drawn on the 12'-0" × 40'-0" pod (11'-0" × 39'-0" interior) · ~430 sq ft · placed beside the clinic
What it adds
- 2 chamber bays (standard), each approx. 8 ft × 12 ft with 36 in clear on all sides
- Compressor and oxygen concentrator room at the far end
- Recovery lounge and reception at the entry
- A 3rd chamber (maximum) needs a second pod: see the capacity check in the notes
Building systems
- Oxygen fire safety (NFPA 99 Ch. 14)
- Dedicated 20–30 A circuits
- Ventilation & cooling
- Plug-and-play utilities
Rooms
| Reception | 55 SF |
| Recovery Lounge | 66 SF |
| Chamber Bays | 264 SF |
| Compressor & O₂ | 44 SF |
Furniture & equipment
- • Hard-shell monoplace hyperbaric chamber, 2.0 ATA (2 standard, 3 maximum)
- • Chamber control panel
- • Air compressor
- • Oxygen concentrator
- • Oxygen monitor
- • Fire extinguisher
- • Recovery recliner
- • Reception desk
- • Guest chair
Preliminary. Confirm with vendor site-planning guide.
Fit check
| Pods required | 1 pod |
|---|---|
| Floor area | ~430 SF |
| Point load | Chamber, approx. 125 kg each |
| Power | Dedicated 20–30 A circuits |
| Shielding | None. O₂ fire safety to NFPA 99 Ch. 14 |
Preliminary. Confirm with vendor site-planning guide.
Equipment values
UnitPreliminary
Hard-shell monoplace chamber, 2.0 ATA
Chamber sizePreliminary
90 cm diameter × 220 cm length (typical)
Chamber weightPreliminary
approx. 125 kg (typical)
ClearancePreliminary
approx. 8 ft × 12 ft per chamber, with 36 in on all sides
CapacityPreliminary
2 chambers standard, 3 maximum (3 need a second pod)
Every value above: Preliminary. Confirm with vendor site-planning guide.
Design notes
- Capacity check: 36 in on all sides of a 90 cm × 220 cm chamber is approx. 9 ft × 13 ft 3 in, a little more than the 8 ft × 12 ft allowance. Two chambers in line with a shared 36 in aisle need approx. 23 ft 5 in and fit the 24 ft bay as drawn
- A 3rd chamber needs approx. 33 ft 7 in of the 39 ft pod, leaving under 6 ft for reception, lounge and the compressor room. Plan 3 chambers across two pods, or move reception and the lounge to another pod
- Oxygen fire safety to NFPA 99 Chapter 14
- Dedicated 20 to 30 A circuits
- Ventilation and cooling for the chamber bays and the compressor room
- Short sessions for cruise guests
Longevity & Destination Wellness: design rules
- Fast set-up and removal
- Plug-and-play utility connections
- Coastal duty: corrosion-resistant finishes
- Coastal duty: hurricane tie-downs