Skip to content
← RESPAND™ platform
Advanced ImagingRP-MRI3 podsNewPreliminaryLicensed service

MRI Suite

1.5T MRI across three joined pods.

MRI Suite floor plan, RP-MRISCHEMATIC · PRELIMINARYRF CAGEREINFORCED FLOOR1.5T MAGNETTABLEQUENCH VENTCONSOLECABINETSCHILLERCORRIDORSCAN ROOM667 SFCORRIDORCORRIDORPATIENT CHANGE77 SFCONTROL110 SFEQUIPMENT132 SFCORRIDOR5-GAUSS LINE (TYPICAL)POD 1POD 2POD 35'-0"29'-0"5'-0"40'-0" exterior36'-0" ext. (3 pods)

Schematic layout (preliminary), drawn on the 12'-0" × 40'-0" pod (11'-0" × 39'-0" interior) × 3 pods · ~1,290 sq ft · dashed walls mark where the next pod connects

What it adds

  • Pods 1 and 2 join side by side with no shared wall, making one scan room approx. 23 ft wide inside
  • Scan room holds the magnet, the RF cage and the 5-gauss zone
  • Pod 3: control room, equipment room (chiller, cabinets) and patient change area
  • Control room shares the long wall with the scan room, with an RF-shielded viewing window

Building systems

  • RF shielding
  • Quench vent
  • Chiller
  • Dedicated power
  • Reinforced floor

Rooms

Scan Room667 SF
Patient Change77 SF
Control110 SF
Equipment132 SF

Furniture & equipment

  • • United Imaging uMR 570 1.5T MRI (70 cm bore)
  • • Patient table
  • • RF cage (Faraday cage) with RF-shielded door and viewing window
  • • Quench vent pipe to the exterior
  • • MRI control console and workstation
  • • Operator chair
  • • Gradient and RF cabinets
  • • Chiller
  • • Power distribution unit
  • • Change cubicles and lockers

Preliminary. Confirm with vendor site-planning guide.

Fit check

Pods required3 pods
Floor area~1,290 SF
Point loadMagnet, approx. 4,000–5,000 kg across pods 1 and 2
PowerDedicated supply; chiller
ShieldingRF cage; magnetic check of adjacent spaces

Preliminary. Confirm with vendor site-planning guide.

Equipment values

  • UnitPreliminary

    United Imaging uMR 570 (1.5T)

  • BorePreliminary

    70 cm

  • Magnet lengthPreliminary

    150 cm

  • Magnet weightPreliminary

    approx. 4,000 to 5,000 kg (typical 1.5T), taken as a concentrated load across the joined pods

Every value above: Preliminary. Confirm with vendor site-planning guide.

Design notes

  • RF shielding: a Faraday cage around the scan room
  • Quench vent from the magnet to the exterior
  • 5-gauss line kept inside the scan room and controlled
  • Magnetic shielding check for adjacent spaces and pods
  • Chiller and dedicated power
  • Plan the delivery route for the magnet into the scan room

Add MRI Suite to your clinic

Configure in 3D and reserve your engineering and production slot.

Open in configurator